Tap Tap to Zoom
Code: S1-9TE-9490534
MPN: 9490534
Sheldon 9490534 Stacking Plate Set for Sheldon Signature HEPA Clean CO2 Incubator is a reliable and dependable addition to the Sheldon Baths, Ovens, Incubators, and Cryogenics Storage family of products. Combining top-notch and uncompromising quality with an affordable price, the Sheldon 9490534 Stacking Plate Set for Sheldon Signature HEPA Clean CO2 Incubator can fulfill your laboratory needs while still offering a great value for the money.![]()
Product Category Description:
This VWR Signature CO2 incubator, with its HEPA filtration system and patented copper housing, draws chamber air through the filter system, removing and destroying airborne microbes and isolated particulates. The copper oxidation kills contaminants. The HEPA filter has an efficiency rating of 99.97% at 0.3µm.
A seamless chamber with radius corners makes it easy to clean. An air jacket around the chamber ensures precise temperature uniformity, reducing the possibility of contamination due to condensation within the chamber. Easily removable shelf system further simplifies the cleaning and decontamination process.
An infrared (IR) CO2 sensor ensures accurate detection and control of CO2 levels in chamber. Unaffected by changes in temperature and humidity, the sensor quickly detects a change in the CO2 concentration as small as 0.1% and compensates accordingly, allowing for rapid CO2 recovery. Humidification is achieved through natural evaporation of distilled water in the humidity pan.
Stacking plate set includes two plates and hardware for securing two incubators together.
| Description:Â | Stacking Plate Set | |
| Unit:Â | Each | |
| Capacity:Â | 0.16 m3 (5.67 cu. ft.) | |
| Interior Dimensions:Â | 49W x 51D x 65H cm | |
| Exterior Dimensions:Â | 65W x 69D x 96H cm | |
| Shelf Dimensions:Â | 45W x 46D cm (173/4 x 181/4") | |
| Electrical:Â | 115V, 50Hz, 700W | |
| Temperature Range: | 8° above ambient to 60°C | |
| Temperature Uniformity: | ±0.35° at 37°C | |
| CO2 Setpoint/Setability: | Digital/0.1°C | |
| CO2 Control Range: | 0–20% | |
| CO2 Control Accuracy: | ±0.1% | |